Sanyo Refrigerator MBR 304DR User Guide

INSTRUCTION MANUAL  
Blood Bank Refrigerator  
MBR-304DR  
MBR-304GR  
MBR-304D  
MBR-304G  
MBR-304G  
 
PRECAUTIONS FOR SAFE OPERATION  
It is imperative that the user complies with this manual as it contains  
important safety advice.  
Items and procedures are described so that you can use this unit correctly and safely.  
If the precautions advised are followed, this will prevent possible injury to the user and  
any other person.  
Precautions are illustrated in the following way:  
WARNING  
Failure to observe WARNING signs could result in a hazard to personnel  
possibly resulting in serious injury or death.  
CAUTION  
Failure to observe CAUTION signs could result in injury to personnel and  
damage to the unit and associated property.  
Symbol shows;  
this symbol means caution.  
this symbol means an action is prohibited.  
this symbol means an instruction must be followed.  
Be sure to keep this manual in a place accessible to users of this unit.  
< Label on the unit >  
This mark is labeled on the cover in which the electrical components of high voltage are  
enclosed to prevent the electric shock.  
The cover should be removed by a qualified engineer or a service personnel only.  
2
 
PRECAUTIONS FOR SAFE OPERATION  
WARNING  
Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to  
rain water.  
Only qualified engineers or service personnel should install the unit. The installation by  
unqualified personnel may cause electric shock or fire.  
Install the unit on a sturdy floor. If the floor is not strong enough or the installation site is not  
adequate, this may result in injury from the unit falling or tipping over.  
Never install the unit in a humid place or a place where it is likely to be splashed by water.  
Deterioration of the insulation may result which could cause current leakage or electric shock.  
Never install the unit in a flammable or volatile location. This may cause explosion or fire.  
Never install the unit where acid or corrosive gases are present as current leakage or electric  
shock may result due to corrosion.  
Use a dedicated power source as indicated on the rating label attached to the unit. Ensure that a  
suitable fuse of adequate amperage is used to allow for compressor start up.  
Remove dust from the power supply plug before inserting in a power source. A dusty plug or  
improper insertion may pose a hazard.  
Use a power supply outlet with ground (earth) to prevent electric shock. If the power supply  
outlet is not grounded, it will be necessary to install a ground by qualified engineers.  
Never ground the unit through a gas pipe, water main, telephone line or lightning rod. Such  
grounding may cause electric shock in the case of an incomplete circuit.  
Do not insert metal objects such as a pin or a wire into any vent, gap or any outlet for inner air  
circulation. This may cause electric shock or injury by accidental contact with moving parts.  
Never store volatile or flammable substances in this unit. This may cause explosion or fire.  
Never store corrosive substances in this unit. This may lead to damage to the inner components  
or electric parts.  
If this unit is to be used for storing poisons, radioactive material or other harmful products,  
ensure that it is in a safe area. Failure to do so may lead to an adverse effect on the health of  
personnel in the area and the local environment. In this case, a request for repair or maintenance  
will necessitate a safety check sheet for maintenance personnel.  
3
 
PRECAUTIONS FOR SAFE OPERATION  
WARNING  
Disconnect the power supply to the unit prior to any repair or maintenance of the unit in order to  
prevent electric shock or injury.  
Ensure you do not inhale or consume medication or aerosols from around the unit at the time of  
maintenance. These may be harmful to your health.  
Never splash water directly onto the unit as this may cause electric shock or short circuit.  
Never disassemble, repair, or modify the unit yourself. Any such work carried out by an  
unauthorized person may result in fire or injury due to a malfunction.  
Disconnect the power supply plug before replacement of the florescent light to prevent electric  
shock.  
Disconnect the power supply plug if there is something wrong with the unit. Continued  
abnormal operation may cause electric shock or fire.  
If the unit is to be stored unused in an unsupervised area for an extended period, ensure that  
children do not have access and that doors cannot be closed completely.  
The disposal of the unit should be undertaken by appropriate personnel. Remove door to  
prevent accidents such as suffocation.  
Prepare a safety check sheet when you request any repair or maintenance for the safety of service  
personnel.  
Do not pull out more than 2 drawers at the same time (MBR-304DR, MBR-304GR) when storing  
or taking out of the stocked material. This may cause injury by falling or tipping of the unit if  
something heavy items are stored.  
CAUTION  
Select a level and sturdy floor for installation. This precaution will prevent the unit from tipping.  
Improper installation may result in water spillage or injury from the unit tipping over.  
Never install the unit in a place where the possibility exists that an object may fall on it. There  
are cooling circuits and electric components in the upper part of this unit. Current leakage or electric  
shock may be caused if an object falls on it.  
4
 
PRECAUTIONS FOR SAFE OPERATION  
CAUTION  
Connect the unit to a power source as indicated on the rating label attached to the unit. Use  
of any other voltage or frequency other than that on the rating label may cause fire or electric shock.  
Fix the shelves securely (MBR-304D, MBR-304G). Incomplete installation may cause injury or  
damage.  
When removing the plug from the power supply outlet, grip the power supply plug, not the cord.  
Pulling the cord may result in electric shock or fire by short circuit.  
Never damage or break the power supply plug or cord. Do not use the supply plug if its cord  
is loose. This may cause fire or electric shock.  
Do not touch any electrical parts such as the power supply plug or any switches with a wet hand.  
This may cause electric shock.  
Do not put a container with water or heavy articles on the unit. Injury may be caused if the  
articles fall. Current leakage or electric shock may result from the introduction of spilled water into  
electrical components.  
Do not climb onto the unit. This may cause damage to the unit and cause it to tip over.  
Hold the handle when closing the door. This will reduce the likelihood of a trapped finger.  
Do not lean on the door. This may cause injury if the unit tips over.  
Never place your hand on the glass or use excessive force on the glass. Intentional force may  
break the glass resulting in injury from the broken glass.  
Do not touch the condenser directly when the filter is removed for cleaning. This may cause injury  
due to hot surface.  
Disconnect the power supply plug before moving the unit. Take care not to damage the power  
cord. A damaged cord may cause electric shock or fire.  
Dispose of any water in the evaporation tray completely before moving the unit. Spilled or  
splashed water may cause current leakage or electric shock.  
Be careful not to tip the unit over during movement to prevent damage or injury.  
Disconnect the power plug when the unit is not used for long periods.  
Do not put the packing plastic bag used in the packaging within reach of children as suffocation  
may result  
5
 
CAUTIONS FOR USAGE  
If the unit is unplugged or the power to the unit is interrupted, do not restart the unit for at least 5  
minutes. This protects the compressor.  
This inner cabinet is refrigerated by the forced circulation of cooled air inside the chamber. Ensure that  
the intake and exhaust vents are not blocked. Adequate space should be provided between the items  
inside the unit to allow air circulation.  
Never store corrosive materials such as acid or alkali where the container is not completely sealed.  
This could possibly lead to corrosion of the evaporator resulting in refrigerant leakage and subsequent  
loss of cooling ability.  
Once the chamber temperature has stabilized, put the items into the chamber in small batches to  
minimize the temperature increase.  
In the case where condensation forms on the front of the glass due to high ambient humidity, wipe it off  
with a soft and dry cloth.  
For the cleaning of the unit, use a cloth containing diluted neutral dishwashing detergent (Undiluted  
detergent may break the plastic parts. For the dilution, follow the instruction enclosed with the  
detergent). When a diluted neutral dishwashing detergent is used to clean the unit, wipe the unit  
thoroughly with a cloth soaked in clean water. Then wipe the unit with a dry cloth to eliminate the  
moisture.  
Do not clean the unit with scrubbing brushes, acid, thinner, solvents powdered soap, cleanser or hot  
water. These agents can scratch the paint or cause it to peel. Plastic and rubber parts can be easily  
damaged by these materials. Especially never use any volatile solvent to clean the plastic or rubber  
parts.  
Fix the shelves securely (MBR-304D, MBR-304G). Place items on the shelves and leave a space  
between the walls of the cabinet and the contents to allow air circulation. Do not place items on the floor  
of the chamber.  
Always close the door firmly. The door check lamp is lit when the door is open. The alarm buzzer  
sounds two minutes after door opening. The buzzer can be cancelled automatically when the door is  
closed.  
Always open and close the door gently. Rough handling of the unit may lead to stocked items tipping  
over, incomplete closing of the door, or damage to the door gasket.  
Do not put any articles on the unit. Blocking of the top will decrease the heat discharge which can  
cause poor cooling and failure of the unit.  
6
 
ENVIRONMENTAL CONDITIONS  
This equipment is designed to be safe at least under the following conditions (based on the IEC 1010-1):  
1. Indoor use;  
2. Altitude up to 2000 m;  
3. Ambient temperature 5oC to 40oC  
4. Maximum relative humidity 80% for temperature up to 31oC decreasing linearly to 50% relative humidity  
at 40oC;  
5. Mains supply voltage fluctuations not to exceed ±10% of the nominal voltage;  
6. Other supply voltage fluctuations as stated by the manufacturer.  
7. Transient overvoltages according to Installation Categories (Overvoltage Categories) II; For mains  
supply the minimum and normal category is II;  
8. Pollution degree 2 in accordance with IEC 664.  
7
 
REFRIGERATOR COMPONENTS  
3
4
1
2
5
6
20  
10  
11  
7
8
12  
13  
21  
14  
15  
9
18  
19  
16  
17  
MBR-304G  
8
 
REFRIGERATOR COMPONENTS  
1. Door switch: To detect the door status (open/close).  
2. Battery switch: ON-OFF switch for built-in battery for power failure alarm. Always turn on this  
switch when the refrigerator is operated. Turn off this switch when the refrigerator is not used for long  
period to protect the battery.  
3. Power switch: Main switch of the unit.  
4. Glow starter: For a fluorescent lamp. It is recommended to replace the glow starter at the time of  
replacement of fluorescent lamp. Refer to page 21 for the replacement.  
5. Control panel: Panel can be opened when upper right corner of the lower part cover is pushed.  
The running status is displayed on this panel and temperature setting is accomplished through this panel.  
Refer to page 10 for details.  
6. Light switch: This switch is used for turning the fluorescent lamp off and on.  
7. Glass window: The window may have condensation in high humidity environment. Wipe off the  
condensation with a soft dry cloth.  
8. Handle: Always hold the handle when opening/closing the door.  
9. Lock: Turn the key counterclockwise through 180o to securely lock the door.  
10. Fluorescent lamp: 15 W lamp. Normally turn off the lamp. See page 20 for the replacement.  
11. Monitor bottle for temperature sensor (2 locations): The bottles has a sensor to display the  
chamber temperature. Fill the bottle with 200ml of 10% glycerol or water before commencing the  
operation. Refer to page 13 for filling the monitor bottle.  
12. Access port: This port allows temperature measurement cables or sensors to enter the chamber  
from outside.  
13. Shelf (MBR-304D, MBR-304G), Drawer (MBR-304DR, MBR-304GR): Items to be stored in the  
chamber must be placed on the shelves or drawers. The maximum storage weight for each shelf or  
drawer is 20 kg. Do not put stored items directly onto the interior floor of the chamber.  
14. Condensation tray: The condensation on the door is accumulated in this tray. Wipe off the water  
occasionally.  
15. Caster: Used when moving the unit. When installing the unit the castors can be raised from the  
ground by using the leveling feet.  
16. Leveling foot: Adjust the height of the leveling foot to install the refrigerator in level. See page 12  
for details.  
17. Automatic temperature recorder: For details, see the instruction manual of the recorder enclosed  
with the unit.  
18. Air exhaust vent: Do not block this vent. Care should be taken not to expose the stored items to  
direct cooled air.  
19. Inner door: Made of acrylic resin. For cleaning, use the diluted neutral dishwashing detergent  
20. Air intake vent: Do not block this vent. If this vent is blocked, temperature regulation will become  
unstable. Do not insert fingers or any substances into the vent because the circulating fan is running  
inside the vent.  
21. Door latch: By turning the latch downward, the door is locked. A padlock can be provided  
separately.  
9
 
REFRIGERATOR COMPONENTS  
Control panel and keypad  
7
8
9
10  
11  
1
2
3
4
5
6
1. Alarm buzzer stop key (BUZZER): To silence the audible alarm, press this key.  
2. Temperature display key (UPPER/LOWER): By pressing this key, the displayed temperature is  
changed; upper monitor bottle temperature, lower monitor bottle temperature, and average temperature  
of upper and lower monitor bottles are displayed in sequence.  
3. Alarm test key (ALARM TEST): To check the operation of alarm buzzer, lamp, and remote alarm  
terminal.  
4. Calibration key (CAL): To calibrate the temperature. This key is used only when the temperature  
calibration is necessary.  
5. Digit shift key (  
): Pressing this key in the CAL mode will cause the changeable digit to shift.  
6. Numerical value shift key (  
): Pressing this key in the CAL mode will cause the numerical value  
to change.  
7. Door check lamp (DOOR): This lamp is lit when the door is ajar or open.  
8. Alarm lamp (ALARM): This lamp blinks during an alarm condition (temperature alarm or power  
failure alarm).  
9, 10. Temperature display monitor lamp (UPPER, LOWER): Lamp 9 is lit when upper monitor bottle  
temperature is displayed. Lamp 10 is lit when lower monitor bottle temperature is displayed.  
Both 9 and 10 are lit when the average temperature of lower and upper monitor bottle temperature is  
displayed.  
11. Digital temperature indicator: This indicator shows the current chamber temperature.  
10  
 
INSTALLATION  
Installation site  
To operate this unit properly and to obtain maximum performance, install the unit in a location with the  
following conditions:  
• A location not subjected to direct sunlight  
Installation in a location subjected to direct sunlight may lead to inadequate cooling.  
• A location with adequate ventilation  
Leave at least 10 cm around the unit for ventilation. Poor ventilation will result in a reduction of the  
refrigeration capacity.  
• A location away from heat generating sources  
Avoid installing the unit near heat-emitting appliances such as gas ranges or stoves. Heat can cause  
inefficient refrigeration.  
• A location with a sturdy and level floor  
WARNING  
Install the unit on a sturdy floor. If the floor is not strong enough or the installation site is not  
adequate, this may result in injury from the unit falling or tipping over.  
Select a level and sturdy floor for installation. This precaution will prevent the unit from tipping.  
Improper installation may result in water spillage or injury from the unit tipping over.  
• A location where objects will not fall on the unit  
CAUTION  
Never install the unit in a place where the possibility exists that an object may fall on it. There are  
cooling circuits and electric components in the upper part of this unit. Current leakage or electric shock  
may be caused if an object falls on it.  
• A location not prone to high humidity  
WARNING  
Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to  
rain water.  
Never install the unit in a humid place or a place where it is likely to be splashed by water.  
Deterioration of the insulation may result which could cause current leakage or electric shock.  
• A location without flammable or corrosive gas  
WARNING  
Never install the unit in a flammable or volatile location. This may cause explosion or fire.  
Never install the unit where acid or corrosive gases are present as current leakage or electric shock  
may result due to corrosion.  
11  
 
INSTALLATION  
Installation  
1. Remove the packaging materials and tapes  
Remove all transportation packaging materials and  
tapes. Open the doors and ventilate the unit. If  
the outside panels are dirty, clean them with a  
diluted neutral dishwashing detergent. (Undiluted  
detergent can damage the plastic components.  
For the dilution, refer to the instruction of the  
detergent.) After the cleaning with the diluted  
detergent, always wipe it off with a wet cloth.  
Then wipe off the panels with a dry cloth.  
2. Adjust the leveling foot  
Leveling foot  
Fig. 1  
Extend the leveling feet by rotating them as shown  
in the figure until they make contact with the floor.  
Ensure the unit is level. (Fig. 1)  
3. Fix the unit  
Two fixtures are attached to the rear of the frame.  
Fix the frame to the wall by attaching a rope or  
chain between the wall and the fixtures.  
4. Install the filter  
Install the condenser filter at the front bottom of the  
cabinet. (Fig. 2)  
Fig. 2  
4. Ground (earth)  
WARNING  
Use a power supply outlet with ground (earth) to prevent electric shock. If the power supply outlet is  
not grounded, it is necessary to install a ground by qualified engineers.  
Never ground the unit through a gas pipe, water main, telephone line or lightning rod. Such  
grounding may cause electric shock in the case of an incomplete circuit.  
12  
 
BEFORE COMMENCING OPERATION  
Preparation of monitor bottle  
Prior to the operation of the unit, fill the monitor  
bottles (upper and lower) with 10% glycerol or  
water using the following procedure:  
Recorder sensor  
(lower bottle of MBR-304G  
and MBR-304GR)  
Temp. sensor  
1. Remove the top and bottom drawer (or shelf).  
2. Remove the monitor bottle cover as shown in  
the figure 1.  
3. Take out the sensor from the monitor bottle.  
4. Remove the fixing screw and take out the  
monitor bottle.  
Cover  
Fig. 1  
5. Remove the cap by loosing it as shown in the  
Fixing screw  
Cap  
figure 2.  
6. Pour 200ml of 10% glycerol or water into the  
monitor bottle.  
Loosen  
7. Cap the monitor bottle and replace the bottle to  
its original position. Then set the sensor to the  
bottle and replace the cover on the bottle.  
Fig. 2  
13  
 
BEFORE COMMENCING OPERATION  
Remote alarm terminal  
The terminal for remote alarm is installed inside the rear  
cover of the unit. The alarm is outputted from this  
terminal.  
The contact rating is DC 30V, 2 A.  
Contact output: At normal condition “Open”  
At abnormal condition “Close”  
Note:  
The contact is “Close” because it is regarded as  
Terminal  
abnormal when the power supply cord is disconnected  
from the outlet or the power switch is turned off, or in  
the event of power failure.  
The remote alarm is in conjunction with the alarm  
buzzer, the alarm can be silenced by pressing the alarm  
buzzer stop key (BUZZER) except for in the case of  
power failure  
14  
 
START-UP OF UNIT  
Follow the procedures for the initial and consequent operations of the unit.  
1. Connect the power cord to the dedicated outlet with appropriate rating.  
2. On start-up, the alarm buzzer sometimes operates. In this case, stop the buzzer by pressing the  
buzzer stop key (BUZZER). The alarm will operate until the temperature monitor bottle sensor goes into  
the range of 4oC±2oC.  
3. Ensure that both monitor bottles have been filled with 10% glycerol or water as outlined on page 13.  
4. The unit has been factory preset to operate at 4oC±1oC. There is no need to set the temperature.  
5. It takes several hours to stabilize the chamber temperature at the set temperature. Once the unit has  
stabilized, ensure the monitor bottle temperature corresponds with the set temperature.  
6. Turn on the fluorescent light switch to check the light. After checking, turn off the switch if the lighting  
is not necessary.  
7. Once the unit operation has been thoroughly checked, begin slowly placing items into the chamber to  
minimize the temperature rise.  
Operation after power failure  
The refrigerator resumes the operation with temperature setting of 4oC after power failure.  
15  
 
OPERATING INSTRUCTIONS  
Storing of items  
Do not disturb the air circulation in the chamber by the stored items. The disturbed air circulation can  
cause the freezing of stored items or degrade of temperature distribution in the chamber.  
Air-intake vent  
Do not block the air-intake by the stored items near the air-intake vent. Take should be taken not to put  
any paper or plastic chips near the vent. They may be vacuumed up and block the vent.  
Air-exhaust vent  
Do not block the air-exhaust by the stored item near the air-exhaust vent. The stored item may be  
frozen if they put nearby the air-exhaust vent.  
Load line  
The stock area is limited to ensure the air flow in the chamber. The labels load lineare attached to the  
right and left side of the chamber. Always put the items within the specified area.  
Note:  
Do not stock any corrosive materials when the container is not sealed completely. Such material causes  
corrosion of inner components or evaporator.  
16  
 
OPERATING INSTRUCTIONS  
Temperature display  
The unit is set at the factory to obtain the proper temperature automatically (4oC±1oC). There is no  
need to set the temperature.  
Table 1. Temperature display  
: Lamp ON : Lamp OFF  
Operation  
Plug-in the power  
supply cord and turn on  
the power switch.  
Press the temperature  
display key.  
Key operated  
Display  
Display mode  
Average monitor bottle  
UPPER  
1
-----  
Average temperature  
temperature  
LOWER  
Upper monitor bottle  
temperature  
UPPER  
LOWER  
UPPER  
LOWER  
UPPER  
LOWER  
UPPER  
LOWER  
2
3
Upper temperature  
Lower temperature  
Average temperature  
Press the temperature  
display key.  
Lower monitor bottle  
temperature  
UPPER  
LOWER  
Press the temperature  
display key.  
Average monitor bottle  
temperature  
UPPER  
LOWER  
4
5
Repeat from step 2.  
Note:  
The displayed temperature represents the temperature sensed by a thermal sensor attached to the  
monitor bottle mounted to the upper and lower portion of the chamber.  
The displayed temperature may not always be at 4oC. Because the chamber probes are representative  
of high and low average temperatures within the chamber.  
Defrost cycles  
There is no need for routine defrosting of the unit as this occurs automatically as follows:  
1. Cycle defrost  
To keep the chamber temperature stable, the refrigeration compressor is cycled on and off. During off”  
period any frost which has accumulated on the evaporator is melted by energizing a defrost heater. This  
will not have any discernible effect on the chamber temperature.  
2. Forced defrost  
When the ambient humidity is high, or a large amount of damp product is being stored inside the chamber,  
there is a possibility that cycle defrost may not be enough to remove all of the frost on the evaporator. In  
this case, a forced defrost cycle can be initiated.  
When the unit is operating under a forced defrost cycle, the current chamber temperature and dFis  
displayed alternately on the digital temperature display.  
Once the forced defrost cycle is completed, normal operation resumes.  
The temperature of blood bag (400mL) in the chamber rises by approximately 1 to 2oC during the  
defrosting.  
17  
 
OPERATING INSTRUCTIONS  
Calibration of temperature sensor  
The temperature sensors are attached to the monitor bottles. Follow the instructions below for proper  
calibration of temperature sensor.  
Recorder sensor  
(lower bottle of MBR-304G  
and MBR-304GR)  
1. Remove the monitor bottle cover as shown in  
Temp. sensor  
the figure.  
2. Remove the sensor to be calibrated from the  
monitor bottle. Please note that MBR-304G and  
MBR-304GR also has a sensor for the temperature  
recorder in the left side of the lower monitor bottle.  
3. Prepare a bottle or beaker with a solution of  
crushed ice and cold water. Agitate the mixture  
thoroughly.  
4. Put the sensor to be calibrated in the mixture.  
5. Select the temperature display by pressing  
temperature display key to show the proper  
temperature. While continuously agitating the ice  
water mixture, monitor the temperature indicated  
on the display of the sensor to be calibrated.  
Cover  
Fixing screw  
6. Wait for the temperature on the display to stabilize. If the display is not 0oC, calibrate using the  
following procedure. If the display does indicate 0oC, proceed to step 7.  
6-1. Press the CAL key for approximately 5 seconds until C00 is displayed.  
6-2. Enter the code of the probe to be calibrated using the numerical value shift key ( ).  
Upper chamber temperature sensor code is C01.  
Lower chamber temperature sensor code is C02.  
6-3. Press the CAL key. (This makes temperature display 0oC)  
This key operation is canceled with the buzzer if the sensor temperature deviates over +/- 1.5oC.  
In this case, consult with Sanyo dealer or service personnel.  
6-4. The digital display will revert back to current chamber temperature. The calibration that was  
entered in the previous step should now have an effect on the temperature display.  
The digital display should now show indicate 0.0±0.2oC, if not, repeat step 6.  
7. Take the sensor out of the ice water solution and wipe with a soft cloth.  
8. Attach the sensor back into the monitor bottle by reversing the procedure in step 1 and 2.  
* For the span adjustment, contact your service personnel.  
18  
 
ALARMS & SAFETY FUNCTIONS  
This unit has the alarms and safety functions shown in Table 2, and also self diagnostic functions.  
Table 2 Alarms and safety functions  
Alarm & Safety  
High temperature  
alarm  
Low temperature  
alarm  
Protection of  
excessive high  
temperature  
Protection of  
excessive low  
temperature  
Situation  
Indication  
Buzzer  
Safety operation  
If the upper or lower temperature sensor  
senses a temperature 6oC.  
If the upper or lower temperature sensor  
senses a temperature 2oC.  
If the high-temperature thermostat  
sensor senses a temperature  
approximately > 10oC.  
Alarm lamp flashes.  
Intermittent tone Remote alarm  
Intermittent tone Remote alarm  
Alarm lamp flashes.  
-----  
Compressor ON.  
Heater OFF.  
-----  
If the alarm sensor senses a  
Compressor ON.  
Heater OFF.  
-----  
-----  
temperature approximately < 2oC.  
When the power to the unit is  
Power failure alarm disconnected or power switch is  
turned off.  
Alarm lamp flashes.  
Door check lamp is lit.  
Intermittent tone Remote alarm.  
Intermittent tone  
Door alarm  
Auto-return  
When the door is ajar or open.  
with 2 minutes  
delay.  
-----  
When there is no key pressing in  
each setting mode for 90 seconds.  
If the temperature sensor of upper  
monitor bottle is disconnected or  
short-circuited.  
Monitor bottle temperature is  
displayed.  
Alarm lamp flashes.  
E1 and monitor bottle temp.  
are displayed alternately.  
Alarm lamp flashes.  
E2 and monitor bottle temp.  
are displayed alternately.  
Alarm lamp flashes.  
E3 and monitor bottle temp.  
are displayed alternately.  
Alarm lamp flashes.  
E4 and monitor bottle temp.  
are displayed alternately.  
Finishing of calibration  
mode.  
-----  
Intermittent tone Remote alarm.  
Intermittent tone Remote alarm.  
Intermittent tone Remote alarm.  
Intermittent tone Remote alarm.  
If the temperature sensor of lower  
monitor bottle is disconnected or  
short-circuited.  
Thermal sensor  
abnormality  
If the defrost sensor is disconnected or  
short-circuited.  
If the over-temperature sensor is  
disconnected or short-circuited.  
Note:  
The remote alarm is silenced by pressing the buzzer stop key (BUZZER) as the remote alarm is operated  
in conjunction with the buzzer, except for the power failure alarm. All alarm functions will auto-reset.  
The buzzer and remote contacts will re-activate if the condition which caused the alarm is not corrected  
within 30 minutes.  
Alarm conditions will remain active for about 24 hours on a fully charged back-up battery in the event of a  
power failure. Two days operation is required to fully charge the battery after initial start-up if the power  
has been off for an extended period.  
The remote alarm contacts will remain active even if the battery is discharged.  
Alarm test procedure  
The alarm function can be checked by the following procedure:  
1. Press alarm test (ALARM/TEST) key.  
2. The audible and visual alarm indicators activate along with the remote alarm. The alarm condition  
remains active for 30 seconds and then terminates automatically. The audible alarm (buzzer) cannot be  
silenced by pressing the buzzer stop key (BUZZER).  
3. Otherwise the alarm test can be terminated manually at any time during the rest by pressing the  
ALARM/TEST key again.  
19  
 
ROUTINE MAINTENANCE  
WARNING  
Always disconnect the power supply to the unit prior to any repair or maintenance of the unit in  
order to prevent electric shock or injury.  
Ensure you do not inhale or consume medication or aerosols from around the unit at the time of  
maintenance. These may be harmful to your health.  
Cleaning of cabinet  
Clean the unit once a month. Regular cleaning keeps the unit looking new.  
Use a dry cloth to wipe off small amounts of dirt on the outside and inside of the unit and all accessories.  
If some of them are dirty, use a cloth containing diluted neutral dishwashing detergent (Undiluted  
detergent may break the plastic parts. For the dilution, follow the instruction enclosed with the  
detergent). When a diluted neutral dishwashing detergent is used, wipe the cabinet or accessories  
thoroughly with a cloth soaked in clean water. Then wipe the cabinet or accessories unit with a dry cloth  
to eliminate the moisture.  
Never pour water onto or into the unit. Doing so can damage the electrical insulation and may cause  
electric shock or short circuit.  
The compressor and other mechanical part are completely sealed. This unit requires absolutely no  
lubrication.  
Remove dust from the power supply plug periodically. A dusty plug or improper insertion may pose a  
hazard.  
Replacement of fluorescent lamp  
The fluorescent lamp is placed horizontally in the upper front of the chamber. Follow the procedure  
below to replace the lamp.  
1. Turn off the fluorescent light switch and disconnect  
the power supply plug.  
2. Hold the cover of the light and pull downward with  
lead wire to remove the lamp from stopper. (Fig. 1)  
3. Remove the water-proof rubber on the left. (Fig.  
2)  
Fig. 1  
4. Remove the socket on the left. (Fig. 2)  
5. Take out the fluorescent lamp gently from the  
protective cover.  
Water-proof rubber  
Socket  
6. After placing a new fluorescent lamp in the  
protective cover, replace the socket and then replace  
the water-proof rubber.  
7. Fix the light with the stoppers.  
Fig. 2  
20  
 
ROUTINE MAINTENANCE  
Replacement of glow starter  
A glow starter is located upper right side of the cabinet..  
1. Disconnect the power supply plug.  
2. Remove the glow starter and replace with a new  
one.  
Glow starter  
3. Replace the water-proof rubber.  
Cleaning of condenser filter  
This unit is provided with the condenser filter at the bottom of the cabinet. Clean the filter once a month  
even since a clogged filter may cause shorter compressor life as well as the poor cooling.  
Clean the filter by the procedure below.  
1. Turn off the power switch.  
2. Take out the condenser filter as shown in the  
figure.  
3. Wash the filter with water and dry it naturally.  
4. Replace the filter.  
DISPOSAL OF UNIT  
WARNING  
If the unit is to be stored unused in an unsupervised area for an extended period ensure that children  
do not have access and doors cannot be closed completely.  
The disposal of the unit should be accomplished by appropriate personnel. Always remove  
doors to prevent accidents such as suffocation.  
This product has an electric battery.  
廢電池  
請回收  
Please recycle it for effective utilization of resources and protection of our environment.  
Thank you for your cooperation.  
Label indication is obliged to comply with Taiwanese battery regulation.  
21  
 
TROUBLESHOOTING  
If the unit malfunctions, check out the following before calling for service.  
If nothing operates even when switched on  
The power is failed.  
The circuit breaker is inactivated.  
The unit is not connected to the power supply.  
The alarm device is activated  
The temperature in the unit does not match the set value. (On start-up)  
The door was kept opened for a long time.  
The high temperature materials (load) were put in the unit.  
In these cases, the alarm is removed automatically by running the unit for several hours.  
When the unit does not get cold enough  
A large amount of items or warm products was put in the unit.  
The air-exhaust vent is blocked up with containers.  
The unit is in direct sunlight.  
The door is frequently opened.  
A heat source is nearby the unit.  
The ambient temperature is too high.  
The door is not securely closed.  
The heat source is put in the unit.  
The door seal is damaged or foreign substance is inserted between door gasket.  
22  
 
SPECIFICATIONS  
Name  
Blood Bank Refrigerator  
MBR-304G MBR-304DR  
Model  
MBR-304D  
MBR-304GR  
External dimensions  
Internal dimensions  
Effective capacity  
Exterior  
W600 x D680 x H1835 (mm)  
W520 x D490 x H1150 (mm)  
304 L  
302 L  
Acrylic finish baked on zinc galvanized steel  
Acrylic finish baked on zinc galvanized steel  
1 door with glass window  
Interior  
Door  
Insulation  
Rigid polyurethane foamed-in place  
Shelf  
Hard steel wire on polyethylene coating  
-----  
5 pcs.  
-----  
Drawer  
Stainless steel, Handle with card holder  
5 pcs.  
Access port  
Left side, Inner diameter 30 mm  
Cooling method  
Compressor  
Forced cool air circulation  
Hermetic type, Output; 175 W  
Fin and tube type  
Condenser  
Evaporator  
Fin and tube type  
Refrigerant  
R-134a  
Temperature controller  
Temperature display  
Alarm  
Electronic control system  
Digital display  
High temperature, Low temperature, Door, Power failure, Sensor  
Allowable contact capacity: 30 VDC, 2 A  
Nonvolatile static memory  
Remote alarm contact  
Memory back-up  
Fluorescent lamp  
Temperature recorder  
Accessories  
White light 15 W x 1 (Glow starter; FG-1P)  
30-day type  
7-day type  
30-day type  
1 set of key  
6 VDC, 1100 mAh (Ni-MH)  
7-day type  
Battery  
Weight  
125 kg  
140 kg  
Note: The contents of this manual are subject to change without notice.  
PERFORMANCE  
+4oC±1oC  
+5 to 35oC  
24 hours (when the battery is full charged)  
Monitor bottle temperature  
Usable ambient temp.  
Power failure alarm  
duration  
Power source  
110V  
60Hz  
220W  
115V  
60Hz  
220W  
220V  
50Hz  
220W  
220 V  
60Hz  
220W  
230 V  
50Hz  
220W  
240 V  
220-240V,  
50Hz  
50Hz  
Rated power consumption  
Noise level  
220W  
220W  
41 dB [A] (background noise; 20 dB)  
1275 kPa  
Maximum pressure  
Note :The unit with CE mark complies with EC directives 89/336/EEC, 93/68/EEC and 73/23/EEC.  
23  
 
CAUTION  
Please fill in this form before servicing.  
Hand over this form to the service engineer to keep for his and your safety.  
Safety check sheet  
Yes  
No  
1. Refrigerator contents :  
Risk of infection:  
Yes  
Yes  
Yes  
No  
No  
No  
Risk of toxicity:  
Risk from radioactive sources:  
(List all potentially hazardous materials that have been stored in this unit.)  
Notes :  
2. Contamination of the unit  
Unit interior  
Yes  
No  
No contamination  
Decontaminated  
Contaminated  
Others:  
Yes  
Yes  
Yes  
No  
No  
No  
3. Instructions for safe repair/maintenance of the unit  
a) The unit is safe to work on  
Yes  
Yes  
No  
No  
b) There is some danger (see below)  
Procedure to be adhered to in order to reduce safety risk indicated in b) below.  
Date :  
Signature :  
Address, Division :  
Telephone :  
Product name :  
Blood Bank Refrigerator  
Model :  
MBR-  
Serial number :  
Date of Installation :  
Please decontaminate the unit yourself before calling the service engineer.  
24  
 
7FB6P101379001  
(9 Sept. 2004)  
SANYO Electric Biomedical Co., Ltd.  
10-15, Hongo 3-Chome  
Bunkyoku, Tokyo 113-8434 Japan  
CEM  
Recycled paper  
Printed in Japan  
 

Radio Shack Clock Radio 63 963 User Guide
RBH Sound Speaker MCS 88 User Guide
RCA Home Theater System RTD325W User Guide
Remington Electric Shaver Smooth Silky shavers User Guide
Russound Speaker LCR7 User Guide
Samsung CRT Television TXN2670WHF User Guide
Samsung Home Theater System HT DS610 User Guide
Samsung Washer Dryer WM1245A User Guide
Sanus Systems TV Mount 6901 002024 User Guide
Sanyo CRT Television AVM 2550S User Guide